Lunaire Photostability Test Chambers
|
|
|
|
|
|



Lunaire Photostability Test Chamber
Larger image
|
|
As an option, Lunaire CEO910W or CEO932W-2 or -4 model chambers
may incorporate photostability test packages for photostability
testing to standards of the International Conference on Harmonisation
(ICH).
Lamp packages are designed with interchangeable fluorescent light
systems that may include:
- Daylight lamps for full spectrum
- Artificial daylight stability testing
- Ultraviolet A lamps for near ultraviolet testing
Features
Light Banks
Full Spectrum Artificial Daylight
- Fluorescent daylight lamps, with
light source at 400-700 nm
- Electronic dimming ballast
- Light intensity range: 300 to 650
foot candles
- Light variation over the product
area (70% of the shelf area) will be
±30% of setpoint
Near UV-Ultraviolet A
- Ultraviolet lamp with a light source of 300-400nm
- Electronic dimming ballast
- Light intensity range: 7 to 25W/M2
- Light variation over the product area (70% of the shelf area)
will be ±30% of setpoint
Both Full Spectrum Artificial Daylight and Near
UV-Ultraviolet A Banks
- Product tray is 12 inches from the bottom of the light bank
assembly
- One lamp package can be provided in Lunaires CEO910W-2
or -4 models
- Up to 3 lamp packages can be provided in Lunaires CEO932W-2
or -4 models
Light Control System
- The flexibility of the light control system is built around
the use of a microprocessor based digital indicator controller
and a Detector Amplifier. This system can be
configured to control either of the two different types of lamp
systems.
- Detector amplifiers are current-to voltage
converter/amplifiers that are compatible with all SED type detectors,
and can be easily calibrated against a standard NIST
traceable detector or light level in
the field or laboratory.
- Light intensity measured by the detector is sent as a current
signal to the detector amplifier, which amplifies it and converts
it to a 0-5VDC signal. The controller configured as a daylight
or ultraviolet controller, processes this signal and displays
actual light intensity in the light bank area.
- Light intensity is controlled with a 0-10VDC output from the
controller to the electronic
dimming ballast to produce and maintain a steady light intensity
throughout the test cycle.
- Set point and actual light intensity are simultaneously displayed
on the controller. Each light system is equipped with an on/off
switch and an indicating light.
- Note: When photostability fixtures
are incorporated on Lunaire chambers, the temperature uniformity
will change to ±2.0°C after stabilization and ±5%
RH on humidified models with the lights on or off.
Other Features:
- Light banks are designed for easy removal for lamp change outs
and are installed using the standard shelf slides. Ballasts are
located on the rear of the chamber for
servicing.
- Lamp bank construction incorporates a reflective material that
offers the same high
reflectance levels for both visible and ultraviolet wavelengths.
- A hinged reflective cover panel on the front of each bank allows
chamber door opening without disruption of light testing.
|
|
|
| |
|
|
|
 |
|
|